Simon PechaI; Heiko BurgerII; Liesa CastroI; Nils GosauIII; Gülsen AtlihanIV; Stephan WillemsIII; Hermann ReichenspurnerI; Samer HakmiI
DOI: 10.21470/1678-9741-2018-0289
ABSTRACT
The Bridge Occlusion Balloon is a compliant balloon,
specifically designed for temporary Superior vena cava occlusion
in case of Superior Vena Cava laceration during lead extraction
procedures.
We here report the first case, using Bridge Occlusion Ballon
for a venous angioplasty in a patient with dysfunctional
pacemaker leads and symptomatic Superior Vena Cava occlusion.
After successful lead extraction, venography was showing a
narrow venous canal. Therefore, venous angioplasty using the
Bridge balloon was performed.
Especially for high-risk lead extraction cases in patients with
Superior Vena Cava stenosis, the Bridge Occlusion Ballon might
be used as a combination of a safety-net in case of Superior Vena
Cava perforation and for Superior Vena Cava angioplasty.
AF = Atrial fibrillation
AV = Atrioventricular
BOB = Bridge Occlusion Balloon
DDD = Dual chamber pacemakers
ICD = Implantable cardioverter defibrillator
OR = Oddis ratio
RV = Right ventricle
SVC = Superior vena cava
TLE = Transvenous lead extraction
INTRODUCTION
Venous stenosis or occlusion is common in patients with multiple- and old ICD- or pacemaker leads[1]. In patients with symptomatic venous stenosis, a class I recommendation for lead extraction is given[2].
In certain cases, even the extraction of all lead material from vascular space does not result in sufficient vascular recanalization. Sometimes, the use of an oversized powered extraction tool might help to create a larger canal through the SVC. However, in cases of long-segment venous stenosis, a high risk for recurrent venous occlusion is given, especially when a re-implantation of leads is necessary. In those patients, a venous angioplasty might be helpful, to allow for free drainage of the venous blood through the SVC, thereby resolving symptoms associated with venous stasis[3]. The BOB (Spectranetics Corporation, Phillips Medical), is a newly designed, compliant balloon, intended for venous occlusion in case of SVC laceration during TLE[4]. However, due to its mechanical properties and specific design for use in the venous system, it might be suitable for venous angioplasty. The device is a compliant, low-pressure balloon. According to its compliance, the risk for venous tear or rupture might be reduced, compared to a rigid high pressure angioplasty balloon. We here report the first case of angioplasty using the BOB during lead extraction.
CASE REPORT
An 86-years old male patient was implanted with a right-sided DDD-pacemaker due to Sick sinus syndrome in 2000 and received a device exchange and new implantation of a RV lead in 2007.
At the time of lead extraction procedure, the patient presented with a third-degree AV-block and paroxysmal AF. Due to atrial fibrillation, the patient was on phenprocoumon, which was maintained during and after the extraction procedure. The indication for lead extraction was SVC occlusion with symptomatic venous stasis and recurrent dizziness, headaches and jugular swelling (Figure 1).
The patient was treated in a hybrid operating room using general anesthesia with cardiopulmonary bypass circuit standby. The RA and RV leads implanted in 2000 were extracted without difficulties using a 14 Fr. 80 Hz Laser sheath including outer sheaths. Afterwards, a venography was performed, showing only a narrow venous recanalization, carrying a high risk for subsequent re-occlusion.
For this purpose, a super stiff wire has been placed into the 14 Fr. Glide light laser sheath. The wire was then subsequently snared and externalized through the right femoral vein. The wire served as a rail for fluoroscopic guided introduction of the BOB.
The balloon was inflated three times at upper, middle and lower SVC level with subsequent volumes of 40ml, 50ml and 60 ml of 20/80 contrast agent/saline mixture.
The following venography confirmed an unimpeded drainage of the contrast medium through the SVC to the right atrium (Figure 2).
The patient was extubated in the OR and went back to regular ward. No clinically significant pulmonary embolism or hemodynamic compromise was observed. The further postoperative course was uneventful, the patient had no further jugular swelling or headache and was discharged without any complications. In the 6-month follow-up period, the patient was doing clinically well without signs of jugular swelling, dizziness or headache.
DISCUSSION
In this case, we have shown for the first time, that the BOB can be successfully used for venous angioplasty.
In case of long-segment venous stenosis, a high risk for recurrence of venous occlusion is given, especially when a re-implantation of leads is necessary. Here, a lead extraction with a powered extraction tool is often not enough, to prevent the patient from recurrent venous stenosis. In those cases, a venous angioplasty might be helpful to resolve the venous stenosis. In some cases, even an angioplasty alone might not be sufficient, especially in the long-term run and some kind of SVC stenting might be necessary[3].
The Bridge occlusion balloon is a newly designed, compliant balloon, that is intended for venous occlusion in cases of SVC laceration during TLE[4-6]. Due to its properties and specific design for the use in the venous system, it might be suitable for venous angioplasty. The device is a compliant, low-pressure balloon and could thereby, minimize the risk for venous tear or rupture, in comparison to a rigid angioplasty balloon. SVC rupture during angioplasty is a rare, but severe complication. Some cases of SVC rupture following angioplasty, have been described in literature, with fatal outcomes, including death. Therefore, the use of a compliant balloon might be favorable to reduce the risk of SVC rupture.
Furthermore, the Bridge occlusion balloon can be prophylactically used in high-risk TLE cases[5,6]. Its prophylactic use has been shown to be safe in short and long-term run. Especially, in cases with SVC stenosis, the BOB might be used as a combination of a safety-net in case of SVC perforation and for SVC angioplasty. Although, there is a high rate of subsequent re-stenosis after venous angioplasty reported in literature[3], the patient had no symptoms of venous stasis in the 6-month follow-up period and was doing clinically well. The fact that the patient was on phenprocoumon due to its atrial fibrillation, might have helped to prevent from re-stenosis. No clear evidence on postoperative anticoagulation after venous angioplasty was found. However, a certain time of anticoagulation treatment seems to be reasonable, to prevent from immediate re-stenosis. Type and timing of anticoagulation needs to be further investigated in larger studies.
REFERENCES
1. Wazni O, Epstein LM, Carrillo RG, Love C, Adler SW, Riggio DW, etal. Lead extraction in the contemporary setting: the LExICon Study: naobservational retrospective study of consecutive laser lead extractions. J AmColl Cardiol. 2010;55(6):579-86.doi:10.1016/j.jacc.2009.08.070.
2. Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI,Birgersdotter-Green UM, Carrillo R, et al. 2017 HRS Expert Consensus Statementon Cardiovascular Implantable Electronic Device Lead Management and Extraction.Heart Rhythm. 2017;14(12):e503-e51.doi:10.1016/j.hrthm.2017.09.001.
3. Brown KT, Getrajdman GI. Balloon dilation of the superior vena cava(SVC) resulting in SVC rupture and pericardial tamponade: a case report andbrief review. Cardiovasc Intervent Radiol. 2005;28(3):372-76.doi:10.1007/s00270-004-0001-0.
4. Clancy JF, Carrillo RG, Sotak R, Ram R, Ryu RK, Kennergren C.Percutaneous occlusion balloon as a bridge to surgery in a swine model ofsuperior vena cava perforation. Heart Rhythm. 2016;13(11):2215-20.doi:10.1016/j.hrthm.2016.06.028. [MedLine]
5. Tsang DC, Azarrafiy R, Pecha S, Reichenspurner H, Carrillo RG, HakmiS. Long-term outcomes of prophylactic placement of an endovascular balloon inthe vena cava for high-risk transvenous lead extractions. Heart Rhythm.2017;14(12):1833-38. doi:10.1016/j.hrthm.2017.08.003. [MedLine]
6. Pecha S, Vogler J, Reichenspurner H, Hakmi S. The bridge occlusionballoon as a safety net in a high-risk transvenous lead extraction procedure.Interact Cardiovasc Thorac Surg. 2018;26(2):360-1.doi:10.1093/icvts/ivx296. [MedLine]
No financial support.
Conflict of Interest: Samer Hakmi and Heiko Burger are Proctors for Spectranetics (Phillips Medical), Simon Pecha received Speaker honoraria from Spectranetics (Phillips Medical).
Authors' roles & responsibilities
SP Conception of the study, manuscript writing; final approval of the version to be published
HB Revising and approval of the manuscript; final approval of the version to be published
LC Manuscript writing; final approval of the version to be published
NG Revising and drafting of the manuscript; final approval of the version to be published
GA Acquisition of data, analysis of data and figures; final approval of the version to be published
SW Revising and approval of the manuscript; final approval of the version to be published
HR Revising and approval of the manuscript; final approval of the version to be published
SH Conception of the study, manuscript writing; final approval of the version to be published
Article receive on Monday, September 17, 2018
Article accepted on Thursday, November 8, 2018